English: chematic drawing of the FM-AFM setup. A very small tip is mounted on an
oscillating cantilever 1 that is in vicinity of a sample. The oscillation of the cantilever
changes upon interaction between the tip and the sample and is detected with a laser beam
2 focussed on the back of the cantilever. The reflected beam travels via mirrors to a position
sensitive detector (PSD) 3 . The signal of the PSD is amplified by a preamplifier. An
amplitude control 4 measures the amplitude A of this signal and a feedback loop compares
it with a setpoint and determines the amplification (dissipation Γ) of the excitation signal
6 for the cantilever which is fed to a shaking piezo. To measure the current resonance
frequency, a phase-locked loop (PLL) 5 is used. Its voltage-controlled oscillator (VCO)
produces the excitation signal 6 for the cantilever. The detected frequency shift ∆f is
passed to another feedback loop 7 that keeps the frequency shift constant by changing the
distance between the tip and the surface (z position) by varying the voltage applied to the
برای به اشتراک گذاشتن – برای کپی، توزیع و انتقال اثر
تلفیق کردن – برای انطباق اثر
تحت شرایط زیر:
انتساب – شما باید اعتبار مربوطه را به دست آورید، پیوندی به مجوز ارائه دهید و نشان دهید که آیا تغییرات ایجاد شدهاند یا خیر. شما ممکن است این کار را به هر روش منطقی انجام دهید، اما نه به هر شیوهای که پیشنهاد میکند که مجوزدهنده از شما یا استفادهتان حمایت کند.
انتشار مشابه – اگر این اثر را تلفیق یا تبدیل میکنید، یا بر پایه آن اثری دیگر خلق میکنید، میبایست مشارکتهای خود را تحت مجوز یکسان یا مشابه با ا اصل آن توزیع کنید.
این پرونده حاوی اطلاعات اضافهایست که احتمالاً دوربین دیجیتال یا پویشگری که در ایجاد یا دیجیتالی کردن آن به کار رفته آن را افزوده است. اگر پرونده از وضعیت ابتداییاش تغییر داده شده باشد آنگاه ممکن است شرح و تفصیلات موجود اطلاعات تصویر را تماماً بازتاب ندهد.